상세 보기
초록
We report on the operational principles and the practical implementation of a near-field scanning microwave microscope system consisting of lambda/4 coaxial resonator with a tunable resonance cavity coupled to a sharp probe tip. The changes of sensitivity and spatial resolution as a function of cavity length are clearly modulated by tuning the resonance cavity. By tuning the resonance cavity, we demonstrate improved sensitivity and spatial resolution better than 4 (mmu) of the near-field images of a YBa2Cu3Oy thin film on a MgO substrate at an operating frequency of f = 1-1.5 GHz. We modelled the lambda/4 coaxial resonator by a LCR circuit and the principles of operation can be explained by using perturbation theory, considering the radius of the probe tip, the quality factor, impedance matching, and the sample-tip distance.
키워드
- 제목
- Development of a near-field scanning microwave microscope using a tunable resonance cavity for high resolution
- 저자
- Kim, J; Kim, MS; Lee, K; Lee, J; Cha, DJ; Friedman, B
- 발행일
- 2003-01
- 유형
- Article
- 권
- 14
- 호
- 1
- 페이지
- 7 ~ 12