Noninvasive in vitro measurement of pig-blood D-glucose by using a microwave cavity sensor

  • Kim, Seungwan
  • Melikyan, Harutyun
  • Kim, Jongchel
  • Babajanyan, Arsen
  • Lee, Jung-Ha
  • 외 3명
Citations

WEB OF SCIENCE

63
Citations

SCOPUS

74

초록

We have developed an electromagnetic microwave cavity sensor based on the resonant frequency shift for real time measurement of the glycemia in pig blood. We could determine the concentration of D-glucose in pig blood in the range of 150-550 mg/dl at the resonance frequency near 4.75 GHz with a bandwidth of 300 MHz. The change in the D-glucose concentration in blood brings microwave reflection coefficient S-11 changes of about 6.26 dB and resonance frequency shifts of about 11.25 MHz due to the electromagnetic interaction between the cavity resonator and the blood filled plastic tube inserted into the cavity. This proposed system provides a unique approach for real time noninvasive and contactless glucose monitoring. (C) 2012 Elsevier Ireland Ltd. All rights reserved.

키워드

NoninvasiveReal time monitoringPig bloodD-GlucoseMicrowave cavity sensorDIELECTRIC-PROPERTIESCELL-MEMBRANETEMPERATUREBIOSENSORSINSULIN
제목
Noninvasive in vitro measurement of pig-blood D-glucose by using a microwave cavity sensor
저자
Kim, SeungwanMelikyan, HarutyunKim, JongchelBabajanyan, ArsenLee, Jung-HaEnkhtur, LkhamsurenFriedman, BarryLee, Kiejin
DOI
10.1016/j.diabres.2012.01.018
발행일
2012-06
유형
Article
저널명
Diabetes Research and Clinical Practice
96
3
페이지
379 ~ 384