Performance evaluation of a PET detector consisting of an LYSO array coupled to a 4x4 array of large-size GAPD for MR compatible imaging

  • Hong, Key Jo
  • Choi, Yong
  • Kang, Jihoon
  • Hu, Wei
  • Jung, Jin Ho
  • 외 3명
Citations

WEB OF SCIENCE

16
Citations

SCOPUS

21

초록

We examined a PET detector consisting of an LYSO array coupled to a 4 x 4 array of large-size Geiger-mode avalanche photodiode (GAPD). The GAPD coupled to 3 mm x 3 mm x 20 mm LYSO pixel crystal has been investigated for possible use as an MR-compatible PET photosensor. Primary characteristics of a PET detector, such as energy resolution and coincidence timing resolution were measured. Gain variation, count uniformity, and count estimation error of 4 x 4 array of LYSO-GAPD were measured to evaluate the performance parameters relevant for PET imaging. The energy resolution and coincidence timing resolution with 511 keV gamma rays were 18.5 +/- 0.7% and 1.6 ns, respectively. The gain variation, count uniformity for all 16 channels were 1.3:1 and 1.3:1, respectively. The count estimation error between adjacent channels measured with an LYSO connected to a GAPD pixel was negligible (0.24 +/- 0.04%). Long-term stability results show that there was no significant change in the photopeak position, energy resolution and count rate for 20 days. Cable lengths up to 300 cm, used between the GAPD and preamplifier, did not affect photopeak position and energy resolution. The performance of the LYSO-GAPD detector inside the MRI exhibited no significant change compared to that measured outside the MRI. The MR images acquired with and without the operating LYSO-GAPD detector located on top of the RF coil showed no considerable degradation in image quality. These results demonstrate the feasibility of using the LYSO-GAPD detector as PET photosensors, which could be used for MR compatible PET development.

키워드

Photon detectors for UV, visible and IR photons (solid-state) (PIN diodes, APDs, Si-PMTs, G-APDs, CCDs, EBCCDs, EMCCDs etc)Gamma camera, SPECT, PET PET/CT, coronary CT angiography (CTA)Multi-modality systemsPOSITRON-EMISSION-TOMOGRAPHYSIMULTANEOUS ACQUISITIONSILICON PHOTOMULTIPLIERAVALANCHE PHOTODIODESMULTICHANNEL PMTSSCANNERREADOUTIMAGESMULTISLICEDEPTH
제목
Performance evaluation of a PET detector consisting of an LYSO array coupled to a 4x4 array of large-size GAPD for MR compatible imaging
저자
Hong, Key JoChoi, YongKang, JihoonHu, WeiJung, Jin HoMin, Byung JunChung, Yong HyunJackson, Carl
DOI
10.1088/1748-0221/6/05/P05012
발행일
2011-05
유형
Article
저널명
Journal of Instrumentation
6