Silicon micromachined infrared thin-layer cell for in situ spectroelectrochemical analysis of aqueous and nonaqueous solvent system

Citations

WEB OF SCIENCE

6
Citations

SCOPUS

6

초록

This article reports in situ FTIR monitoring of electrochemical reactions using a silicon-based thin-layer cell realized by micromachining fabrication technologies. ne proposed device contains a silicon micromachined cavity cell integrated with a gold working electrode, a counter electrode, and a Ag/AgCl reference electrode. Electrochemical, spectroscopic, and spectroelectrochemical characteristics of the fabricated cell have been experimentally measured and exhibited a comparable performance to commercialized CaF2 thin-layer cells. In situ spectroelectrochemical characteristics have been successfully tested for both 2 mM K3Fe(CN)(6) in an aqueous solution and 2 mM ferrocene in a dichloromethane solution using the fabricated thin-layer cell.

키워드

thin-layer cellspectroelectrochemical analysisFTIRMEMSelectrochemical cellELECTROCHEMICAL-CELLELECTRODESREDUCTIONSPECTROSCOPYCOMBINATIONBILIRUBIN
제목
Silicon micromachined infrared thin-layer cell for in situ spectroelectrochemical analysis of aqueous and nonaqueous solvent system
저자
Yun, KSJoo, SKim, HJKwak, JYoon, E
DOI
10.1002/elan.200403202
발행일
2005-06
유형
Article
저널명
Electroanalysis
17
11
페이지
959 ~ 964