Real-Time Noninvasive Measurement of Glucose Concentration Using a Microwave Biosensor

  • Bababjanyan, Arsen
  • Melikyan, Harutyun
  • Kim, Seungwan
  • Kim, Jongchel
  • Lee, Kiejin
  • 외 1명
Citations

WEB OF SCIENCE

47
Citations

SCOPUS

54

초록

We measured the glucose concentration by using the real-time electromagnetic interaction between probe-tip and glucose solution using a microwave biosensor. The microwave biosensor, consisting of a dielectric resonator coupled to the probe-tip, allows observation of the small variation of the glucose concentration changes in the ranges of 0-300 mg/ml by measuring the microwave reflection coefficient S-11. We could observe the concentration of glucose with a detectable resolution up to 1 mg/ml at an operating frequency of about f = 4-5GHz. The change of the glucose concentration is directly related to the change of the reflection coefficient due to the electromagnetic interaction between the microwave resonator and the glucose solution. The operational principal is explained by the plane-wave solution model. The measured signal-to-noise ratio was about 37 dB, and the minimum detectible signal was about 0.003 dB/(mg/ml). A glucose biosensor using a microwave resonator with probe provides a unique approach for glucose real-time monitoring.

키워드

SHEET RESISTANCEDISTANCE CONTROLMICROSCOPE
제목
Real-Time Noninvasive Measurement of Glucose Concentration Using a Microwave Biosensor
저자
Bababjanyan, ArsenMelikyan, HarutyunKim, SeungwanKim, JongchelLee, KiejinFriedman, Barry
DOI
10.1155/2010/452163
발행일
2010
유형
Article
저널명
Journal of Sensors
2010