INVESTIGATION ON MECHANICAL AND BIO-MECHANICAL PERFORMANCE OF A CENTRIFUGAL BLOOD PUMP

  • 장민욱
  • M. Moshfeghi
  • 허남건
  • 강성원
  • 김원정
  • 외 1명

초록

Blood pump analysis process includes both mechanical and bio-mechanical aspects. Since a blood pump is a mechanical device, it has to be mechanically efficient. On the other hand, blood pumps function is sensitively related to the blood recirculation; hence, bio-factors such as hemolysis and thrombosis become important. This paper numerically investigates the mechanical and bio-mechanical performances of the Rotaflow in the extracorporeal membrane oxygenation(ECMO), Ventricular Assist Device(VAD), and full-load conditions. The operational conditions are defined as(400[mmHg], 5[L/min.]), (100[mmHg], 3[L/min.]), and (600[mmHg], 10[L/min.]) for ECMO, VAD, and full-load conditions, respectively. The results are presented and analyzed from the mechanical aspect via performance curves, and from bio-mechanical aspect via focusing on hemolytic characteristics. Regions of top and bottom cavities show recirculation in both ECMO and VAD condtions. In addition, Eulerian-based calculation of modified index of hemolysis(MIH) has been investigated. The results demonstrate that the VAD condition has the least risk of hemolysis among the others, while the full-load condition has the highest risk.

키워드

Centrifugal Blood PumpCFDMechanical PerformanceECMOVADMIH
제목
INVESTIGATION ON MECHANICAL AND BIO-MECHANICAL PERFORMANCE OF A CENTRIFUGAL BLOOD PUMP
저자
장민욱M. Moshfeghi허남건강성원김원정강신형
DOI
10.6112/kscfe.2015.20.2.088
발행일
2015-06
저널명
한국전산유체공학회지
20
2
페이지
88 ~ 95