Mechanical Effects of Cochlear Implant on Acoustic Hearing

Citations

WEB OF SCIENCE

3
Citations

SCOPUS

3

초록

Residual hearing loss in cochlear implant users is investigated using the mechanical-human-cochlear model. Hearing loss due to stiffening of the round window increases significantly as input frequencies decrease from 3 kHz to 1 kHz but remains constant at lower frequencies, whereas loss due to the presence of an electrode insert becomes significantly higher at lower frequencies (< 1 kHz). The latter also shifts the characteristic frequency map toward the basal end of the cochlea. In the region away from the end of the electrode insert, cochlear function recovers, but the user still suffers from hearing loss caused by round window stiffening.

키워드

Speech recognitionhuman cochleacochlear implantresidual hearingbasilar membraneBASILAR-MEMBRANEINTRACOCHLEAR PRESSUREELECTRODE INSERTIONMODELAMPLIFICATIONDEPTHMORPHOLOGY
제목
Mechanical Effects of Cochlear Implant on Acoustic Hearing
저자
Chan, Wei XuanYoon, Yong-JinShin, Choongsoo S.Kim, Namkeun
DOI
10.1109/TBME.2018.2876402
발행일
2019-06
유형
Article
저널명
IEEE Transactions on Biomedical Engineering
66
6
페이지
1609 ~ 1617