상세 보기
A System-on-Chip Solution for Point-of-Care Ultrasound Imaging Systems: Architecture and ASIC Implementation
- Kang, Jeeun;
- Yoon, Changhan;
- Lee, Jaejin;
- Kye, Sang-Bum;
- Lee, Yongbae;
- ... Yoo, Yangmo;
- 외 3명
WEB OF SCIENCE
47SCOPUS
56초록
In this paper, we present a novel system-on-chip (SOC) solution for a portable ultrasound imaging system (PUS) for point-of-care applications. The PUS-SOC includes all of the signal processing modules (i.e., the transmit and dynamic receive beamformer modules, mid- and back-end processors, and color Doppler processors) as well as an efficient architecture for hardware-based imaging methods (e.g., dynamic delay calculation, multi-beamforming, and coded excitation and compression). The PUS-SOC was fabricated using a UMC 130-nm NAND process and has 16.8 GFLOPS of computing power with a total equivalent gate count of 12.1 million, which is comparable to a Pentium-4 CPU. The size and power consumption of the PUS-SOC are 27 x 27 mm(2) and 1.2 W, respectively. Based on the PUS-SOC, a prototype hand-held US imaging system was implemented. Phantom experiments demonstrated that the PUS-SOC can provide appropriate image quality for point-of-care applications with a compact PDA size (200 x 120 x 45 mm(3)) and 3 hours of battery life.
키워드
- 제목
- A System-on-Chip Solution for Point-of-Care Ultrasound Imaging Systems: Architecture and ASIC Implementation
- 저자
- Kang, Jeeun; Yoon, Changhan; Lee, Jaejin; Kye, Sang-Bum; Lee, Yongbae; Chang, Jin Ho; Kim, Gi-Duck; Yoo, Yangmo; Song, Tai-kyong
- 발행일
- 2016-04
- 유형
- Article
- 권
- 10
- 호
- 2
- 페이지
- 412 ~ 423