We characterized a vertically configured electrical detection system that used scanning tunneling microscopy (STM) to detect antigen-antibody binding. This technique could be used to easily construct a multiple measurement system in a protein chip. We utilized immunocomplexes comprised of our model protein, prostate specific antigen (PSA), corresponding antibody fragments, and gold nanoparticle-antibody conjugates. The electrical tunneling current between the STM tip and these complexes exhibited a peaklike pulse, the frequency of which depended on the surface density of the bound complexes. We could therefore quantitatively measure PSA concentrations as low as 10 fg/mL using periodogram analysis of this peak frequency. (C) 2008 American Institute of Physics.