A New Fast Logarithm Algorithm Using Advanced Exponent Bit Extraction for Software-Based Ultrasound Imaging Systems

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

Ultrasound B-mode imaging provides anatomical images of the body with a high resolution and frame rate. Recently, to improve its flexibility, most ultrasound signal and image processing modules in modern ultrasound B-mode imaging systems have been implemented in software. In a software-based B-mode imaging system, an efficient processing technique for calculating a logarithm instruction is required to support its high computational burden. In this paper, we present a new method to efficiently implement a logarithm operation based on exponent bit extraction. In the proposed method, the exponent bit field is first extracted and then some algebraic operations are applied to improve its precision. To evaluate the performance of the proposed method, the peak signal-to-noise ratio (PSNR) and the execution time were measured. The proposed efficient logarithm operation method substantially reduced the execution time, i.e., eight times, compared to direct computation while providing a PSNR of over 50 dB. These results indicate that the proposed efficient logarithm computation method can be used for lowering the computational burden in software-based ultrasound B-mode ultrasound imaging systems while improving or maintaining the image quality.

키워드

ultrasound B-mode imagingfloat point logarithmsoftware-based ultrasound imaging systemssingle instruction multiple data
제목
A New Fast Logarithm Algorithm Using Advanced Exponent Bit Extraction for Software-Based Ultrasound Imaging Systems
저자
Park, SeongjunYoo, Yangmo
DOI
10.3390/electronics12010170
발행일
2023-01
유형
Article
저널명
Electronics (Basel)
12
1