A microfluidic chip for measurement of biomolecules using a microbead-based quantum dot fluorescence assay

Citations

WEB OF SCIENCE

29
Citations

SCOPUS

27

초록

This paper describes a custom-designed microfluidic chip for sensitive detection of antibody using quantum dot fluorescence in a microbead-based assay. The microfluidic chip is designed to isolate a single microbead where the binding reaction of antibodies happens on the surface. The microfluidic chip is fabricated on a glass substrate using transparent silicone elastomer, PDMS, for easy access to monitoring and flexible gate operations to capture the single microbead. For antibody detection, a sequence of functionalized assay has been performed in the fabricated chip including: capturing of microbeads; antibody injection into a micro-chamber; quantum dot injection; and fluorescence detection. Various concentrations of antibody have been tested and have shown good agreement with correlated fluorescent intensities.

키워드

antibodymicrobeadquantum dotmicrofluidic chipMEMSBEADSPROTEIN
제목
A microfluidic chip for measurement of biomolecules using a microbead-based quantum dot fluorescence assay
저자
Yun, Kwang-SeokLee, DohoonKim, Hak-SungYoon, Euisik
DOI
10.1088/0957-0233/17/12/S10
발행일
2006-12
유형
Article
저널명
Measurement Science and Technology
17
12
페이지
3178 ~ 3183