상세 보기
Assembly of laccase over platinum oxide surface and application as an amperometric biosensor
- Quan, D;
- Kim, Y;
- Yoon, KB;
- Shin, W
WEB OF SCIENCE
41SCOPUS
43초록
Laccase could be successfully assembled on an amine-derivatized platinum electrode by glutaraldehyde coupling. The enzyme layer formed on the surface does not communicate electron directly with the electrode, but the enzymatic activity of the surface could be followed by electrochemical detection of enzymatically oxidized products. The well-known laccase substrates, ABTS (2,2'-azinobis(3-ethylbenzothiazoline-6-sulfonic acid)) and PPD (p-phenylenediamine) were used. ABTS can be detected down to 0.5 muM with linear response up to 15 muM and current sensitivity of 75 nA/muM. PPD showed better response with detection limit of 0.05 muM, linear response up to 20 muM, and current sensitivity of 340 nA/muM with the same electrode. The sensor responses fit well to the Michaelis-Menten equation and apparent K-M values are 0.16 mM for ABTS and 0.055 mM for PPD, which show the enzymatic reaction is the rate-determining step. The laccase electrode we developed is very stable and more than 80% of initial activity was still maintained after 2 months of uses.
키워드
- 제목
- Assembly of laccase over platinum oxide surface and application as an amperometric biosensor
- 저자
- Quan, D; Kim, Y; Yoon, KB; Shin, W
- 발행일
- 2002-03
- 유형
- Article
- 권
- 23
- 호
- 3
- 페이지
- 385 ~ 390