Iterative Simultaneous Attenuation Correction and Activity Estimation for Time-of-Flight Positron Emission Tomography

  • Lee, Nam-Yong
  • Kim, You Suk
  • Lim, Hyun Keong
  • Choi, Yong
Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

An accurate quantitative estimation of the radioactive tracer distribution in positron emission tomography (PET) requires correction of the observed emission data for the attenuation factors. For this purpose, transmission data, provided by X-ray computerized tomography (CT) or magnetic resonance imaging (MRI), are often used to measure the attenuation rate in observed PET data. In this paper, we propose an iterative simultaneous attenuation correction and activity estimation method without transmission data. In each iteration, the proposed method disperses the attenuation effect in the time-of-flight (TOF) PET sinogram in each line-of-response to the current iterates for the activity estimate, then uses that iterate to update the attenuation rate in the TOF PET sinogram, and finally use that updated attenuation rate in computing the next iterates. The simulation result shows that the proposed method corrects attenuation artifact accurately and produces activity estimates comparable to reconstructed ones from attenuation-free data.

키워드

Time-of-flightPETAttenuation correctionIMAGE-RECONSTRUCTIONPET RECONSTRUCTIONPHOSWICH DETECTORHIGH-RESOLUTIONLSODEPTHOPTIMIZATIONSIMULATIONCHALLENGESALGORITHM
제목
Iterative Simultaneous Attenuation Correction and Activity Estimation for Time-of-Flight Positron Emission Tomography
저자
Lee, Nam-YongKim, You SukLim, Hyun KeongChoi, Yong
DOI
10.3938/jkps.77.689
발행일
2020-10
유형
Article
저널명
Journal of the Korean Physical Society
77
8
페이지
689 ~ 699