Development of Triglyceride Certified Reference Materials in Human Frozen Serum Using Isotope Dilution-Liquid Chromatography-Tandem Mass Spectrometry

  • Lee, Sun Young
  • Kim, Kwonseong
  • Kwon, Hyun Soo
  • Kim, Young Eun
  • Kim, Ki Young
  • ... Oh, Han Bin
  • 외 4명
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

Certified reference materials (CRMs), KRISS CRM 111-01-010 and 111-01-011, were first developed in Korea to provide SI-traceability in the measurement of triglycerides (TGs) in human frozen serum. For absolute TG quantification in KRISS CRMs, isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) was used. Both certification and short-term stability assessment of TG in KRISS CRMs were performed for each of 10 and eight selected units. The TG certified values and expanded uncertainties in KRISS CRMs were as follows: 0.608 +/- 0.025 mmol/kg for CRM 111-01-010 and 1.825 +/- 0.072 mmol/kg for CRM 111-01-011. The short-term TG stability results were not significantly changed relative to the certified values, and therefore both KRISS CRMs were stable during the monitoring period at each temperature. Based on our experiments, the developed KRISS CRMs can provide accurate, precise, and reproducible TG measurements in human frozen serum for clinical laboratories with SI-traceability.

키워드

TriglycerideHuman frozen serumIsotope-dilution mass spectrometryCertified reference materialDEFINITIVE METHODTOTAL GLYCERIDESCHOLESTEROL
제목
Development of Triglyceride Certified Reference Materials in Human Frozen Serum Using Isotope Dilution-Liquid Chromatography-Tandem Mass Spectrometry
저자
Lee, Sun YoungKim, KwonseongKwon, Hyun SooKim, Young EunKim, Ki YoungJeong, Ji-SeonOh, Han BinKim, Tae YoungHong, JongkiKang, Dukjin
DOI
10.1002/bkcs.11703
발행일
2019-05
유형
Article
저널명
Bulletin of the Korean Chemical Society
40
5
페이지
418 ~ 423